Arif
Saltuk Ltd |
 |
|
|
Services |
|
Consulting |
|
Viscometer
Calibration |
|
Repair
& Maintenance |
 |
|
|
|
|
|
 |
 |
CAP 1000/2000
High Shear Rate Viscometers
The CAP Viscometer was originally designed to provide a high shear rate test for paints and coatings that simulates material behavior in rolling, brushing, and spraying applications. The versatility of the CAP design makes it a practical tool for any QC or R&D lab requiring quick easy testing of materials, regardless, of application, at high shear rates.
Applications include paints, resins, foods, cosmetics, pharmaceuticals, dispersions, coatings, and inks.
- Shear rates from 166 sec-1 to 26,600 sec-1
- Variable shear rate (CAP 2000)
- Cone and plate geometry
- Small sample size (less than 1 ml)
- Automatic cone gap positioning
- Automatic viscosity range calibration
- Rapid sample temperature control on plate surface
- Digital settings and display (CGS or SI units)
- PC control with RS-232 serial interface (CAP 2000) using Brookfield Capcalc software
- Parallel printer interface to document data
- Timed test capability
- Sample number identification
- Easy test set up and clean up
- Rapid test performance
- Two temperature ranges: 5°C to 75°C OR 50°C to 200°C
- NIST traceable viscosity standards available
- Choice of one cone spindle supplied with viscometer (extras available)
CAP1000 is a single speed viscometer:
- 900 rpm at 60 Hz or 750 rpm at 50 Hz
CAP2000 is a variable speed viscometer:
- 50 to 2000 rpm in 1 rpm steps
|
Cone Spindle Range Table
Cone Spindle No.
|
Cone Angle
|
Shear Rate (sec-1)
|
Full-Scale Viscosity Range (Poise)
|
Sample Size (µl)
|
CAP 1000 at 750 RPM
|
CAP 1000 at 900 RPM
|
CAP 2000 at 50-2000 RPM
|
01
02
03
04
05
06
|
0.45°
0.45°
0.45°
1.8°
1.8°
1.8°
|
13.3N
13.3N
13.3N
3.3N
3.3N
3.3N
|
.25-.25
.5-5.0
1.0-10.0
2.0-20.0
4.0-40.0
10.0-100.0
|
0.21-2.08
0.42-4.17
0.83-8.33
1.67-16.7
3.33-33.3
8.33-83.3
|
0.1-37.5
0.2-75.0
0.4-150.0
0.8-300.0
1.6-600.0
3.8-1500.0
|
67
38
24
134
67
30
|
N=RPM
|
Standard Fluids
|
|